Product News

False "Air in line" alarms with Baxter Flo-Gard IV administration sets.

In July 1997, Baxter sent out a product information notice affecting Flo-Gard IV administration sets manufactured since January 1997, with the following codes:

A change of supplier of the PVC resin has lead to a small change in surface finish and stiffness of the tube on some sets. Corrective action has been taken by Baxter to eliminate this problem, and new batches will be available at the end of July.

In the meantime, Baxter have made available a procedure to allow the current administration sets to be used. The procedure is available from Baxter. In the UK, contact

Myles Darbyshire, Tel: 01635-206411.

The Medical Devices Agency is aware of these reports and the action Baxter have taken.